High resolution beam imaging systems - BIS

High-resolution beam imaging is made possible with the Colutron Beam Imaging Systems (BIS).   The BIS can be used to measure two or three dimensional intensity distributions of ion, electron and neutral beams.   The BIS can also be used to image X-rays for applications such as pinhole imaging and spectroscopy.
Images are created using a microchannel plate (MCP) and phosphor screen.   The phosphor is uniformly deposited onto a coherent fiberoptic (FO) substrate so that the image can be optically transmitted to the outside of a vacuum system for analysis using a FO conduit.
model BIS-3135 with BIS-1-PH probe head
The BIS systems are available in many different configurations with several options for image readout to best fit a customer's particular application and budget.
model BIS-4135 (Remote System) model BIS-4105 (Standard System)

Features:
Available with single or dual MCPs
Microsphere plates (MSP) also available
Spatial resolution down to 25µm
Interchangeable probe heads for changing imaging area, gain and/or resolution
Digital readout CID or CCD cameras
CID/CCD cameras available with fiberoptic faceplate for direct fiberoptic coupling to probe head
Available with flexible fiberoptic cable for remote imaging applications
Phosphor screen available aluminized (standard) or Indium-Tin Oxide (ITO)
Front mounting grids available for electron suppression, beam attenuation, beam acceleration
UHV compatible
Microchannel plate performance enhancement options available

Applications:
High Resolution Single Particle Position Sensitive Detection
Remote Beam Line Diagnostics
Real Time Image Analysis
Remote Beam Profile Analysis (beam tuning)
Mass Spectrometry (mass and dispersion determination)
Low Energy and Low Intensity Beam Imaging
X-ray Spectroscopy
Field Ion Microscopy

4 keV OH+ ion beam ( 0.86mm FWHM ) made with model G-1 ion gun system.   model BIS-1035 beam imaging system, and IPS-1 image processing system.

Beam Imaging System (BIS)-General Description
Probe Heads
Fiberoptic Conduits
CCD and CID Camera Options
BIS Ordering Information
Beam Imaging System Dimensions
Beam Imaging System Specifications

BEAM IMAGING SYSTEM - (BIS)
General Description
The BIS systems consist of three basic components: 1) the BIS probe head which houses the MCP and phosphor screen; 2) a fiberoptic conduit which optically transmits the image from the probe head to a fiberoptic vacuum feedthrough; and, 3) a camera system for digital readout of the image.   These components are described in detail below.
  PROBE HEAD
The BIS probe head creates an image of a beam by converting the incoming beam to electrons by secondary electron emission from the probe head microchannel plate (MCP).   The electrons are then accelerated onto a phosphor screen with enough energy to excite the phosphor and create an image.   The phosphor is deposited uniformly across a coherent fiberoptic (FO) disc or taper.   The BIS-1 systems use a 1:1 magnification FO disc whereas, in order to image larger areas, the BIS-2, BIS-3 and BIS-4 systems use FO tapers with corresponding magnifications of 2x, 3x, and 4x.   Optional microsphere MSP plates are also available in a single or dual stack.   These plates have higher gain per stage, no ion feedback, and less severe vacuum requirements than MCPs, but have poorer spatial resolution (see MSP - MCP Comparison Chart and Imaging System Specifications .   All BIS probe heads are interchangeable, allowing the user to change image resolution and image area by simply switching the probe head.

Probe Head Specifications
Probe Head model number image area approximate spatial resolution
   BIS-1-PH    17mm diam.    25µm - 35µm
   BIS-2-PH    25.4mm diam.    50µm - 70µm
   BIS-3-PH    38.4mm diam.    75µm - 100µm
   BIS-4-PH    ~ 44.45mm diam.    150µm

  PROBE HEAD - OPTIONAL CONFIGURATIONS
Probe heads are available with a single MCP (standard) or dual MCP (chevron) configuration for increased gain.   Also available are microsphere plates (MSP) in single or dual stack.   These plates have higher gain per stage and less severe vacuum requirements, but have poorer spatial resolution (see Beam Imaging Specifications on page 27).
The Phosphor Screens come standard with P-20 phosphor and conductive aluminum coating.   Other phosphors are available upon special request, as well as an Indium-Tin-Oxide (ITO) coating.   ITO screens are generally used when lower electron accelerating voltages are required, however the quantum yield may be lower than that of aluminized screens.
The probe heads also allow for mounting a variety of apertures, slits, lenses, pinholes, filters and grids.   All probe heads have holes for mounting 1/8" (3mm) diameter ceramic mounting rods.   Colutron offers a small selection of grids that can be mounted in front of the probe head for applications such as electron suppression, beam attenuation, and beam acceleration.   A 4-terminal high voltage feedthrough on 2 3/4" (DN35) conflat flange is provided with the BIS system for the phosphor screen and MCP voltage application.
The input microchannel plate is also available with performance enhancement coatings to enhance the detection efficiency for certain types of radiation.   The table below outlines the available coatings.

coating
type
Radiation type
enhanced
Energy Range coating
thickness (Å)
coating extension
(Channel Diameters)
CuI    UV 200 ~ 2,000Å    3,000    1.5
CsI    UV 200 ~ 2,000Å    3,000    1.5
MgFl2    UV 350 ~ 1,200Å    3,000    1.5
KBr    X-Ray 0.2 ~ 9 keV    3,000    1.5

  FIBEROPTIC CONDUITS
Images are transmitted optically from the probe head through one of two types of FO conduits which both include a 2·3/4" (DN35) conflat Ò FO vacuum viewport enabling imaging on the outside of the vacuum chamber CID camera options below).   Conduits consist of either a flexible 60cm long, 10mm x 8mm rectangular FO cable (Remote System) or 19mm diameter solid FO rod (Standard System) which is available in various lengths.   Both conduit types attach at one end to the BIS probe head, and to the FO viewport at the other end.   The remote systems with flexible FO cable are especially useful for imaging particle beams deep within the vacuum chamber where the beam cannot be brought to a vacuum port for imaging.   The 60cm length is standard, however other lengths are available on request.   The standard systems with solid FO rod are generally used when the beam can be brought close to a vacuum port for imaging.   The FO rod is available in various lengths to allow for greater flexibility when adapting to a customer's particular application.
  CCD and CID camera options
Optional CCD and camera readout systems are available for all BIS systems.   All cameras are available with lens, video monitor and mounting system for imaging the FO viewport.   The CIDTEC 3710D and COHU 4812 cameras are also available with fiberoptic faceplates (FOFP) directly attached to the camera sensor.   The camera sensor is then mounted in a special housing which allows them to directly mate to the FO viewport.   This direct FO coupling approach ensures highest possible resolution and gain.
The camera can be connected to a video monitor in order to observe images in real time, and can also be connected to a video frame grabber making it possible to store and analyze the images later.   Colutron offers the optional Image Processing Systems-IPS which include video monitor, frame grabber, cables and software.

 
  CCD / CID camera specifications
camera CIDTEC 3710D COHU 4812 COHU 2122
  Optical Format   2/3" diagonal   2/3" diagonal   1/2" diagonal
  resolution   755H x 484V   755H x 488V   768H x 494V
  Element Pitch (µm)   12.0 x 13.7   11.5 x 27   8.4 x 9.8
  area (mm)   9.05 x 6.83   8.8 x 6.6   6.4 x 4.8
  FOFP (optional)     4.5 µm pitch   4.5 µm pitch   n/a
  scanning format   RS-170, 2:1 interlace   EIA RS-170   EIA RS-170
  electronic shutter         1/60 - 1/10,000 sec.
  sync. system   int. / ext.   int. / ext.   int. / ext.
  signal to noise (dB)   50   50   > 55
  sensitivity
  (face plate illumination)
  full output: 0.5fc   full output: 0.2 lux   full output: 0.65 lux
  input power (Watts)   8.5 max.   4.2   3.6
  input voltage (Volts)   +12V.dc   +12V.dc   + 15V.dc nominal

BIS Ordering Information
Use the table below to determine the BIS model that fits your application.   For example, a model BIS-3135 decodes as a BIS system with BIS-3-PH probe head, dual MCPs, 60cm long fiberoptic cable with 2·3/4" (DN35) conflatÒ viewport, and fiberoptically coupled CIDTEC 3710D camera.   NOTE: All BIS systems include a four terminal, 2·3/4" conflat (DN35) high voltage feedthrough (Colutron part number PN-706).
BIS-X
Probe Head
X
MCP options
X
FO conduit
X
camera system
1 - BIS-1-PH 0 - single MCP 0 - FO rod ( 2" long )
/ FO viewport
0 - none
2 - BIS-2-PH 1 - dual MCP 1 - FO Rod ( 3" long)
/ FO viewport
1 - COHU 2122/Lens/mount
3 - BIS-3-PH 2 - single MSP 2 - FO Rod ( 4" long )
/ FO viewport
2 - COHU 4812/Lens/mount
4 - BIS-4-PH 3 - dual MSP 3 - FO cable ( 60cm )
/ FO viewport
3 - CIDTEC 3710D/Lens/mount
  4 - none ( phosphor only )   4 - COHU 4812/FOFP/mount
  5 - special coating ( specify type )   5 - CIDTEC 3710D/FOFP/mount

REMOTE SYSTEM DIMENSIONS
The dimensions for the BIS system with optional FO cable and CIDTEC camera with FOFP are shown below.
Remote beam imaging system diagram

DIMENSIONS ( Remote Beam Imaging Systems )
  models * BIS-1X35
BIS-2X35 BIS-3X35 BIS-4X35
  dimensions (mm)
  A     83.5   83.5   83.5   83.5
  B     68.3   68.3   68.3   68.3
  C     108.0   108.0   108.0   108.0
  D     63.5   76.2   88.9   101.5
  E     24.8   43.3   56.3   70.09
  F     12.7   12.7   12.7   12.7
  G     566.7   566.7   566.7   566.7
  H     4.7   4.7   4.7   4.7
  Cable end-tip
  O.D. (
f )
    34.3   34.3   34.3   34.3
* note: The model number X descriptor determines the MCP option


STANDARD BEAM IMAGING SYSTEM DIAGRAM
DIMENSIONS ( standard beam imaging systems )
  models * BIS-1X05 BIS-1X15 BIS-1X25 BIS-2X05 BIS-2X15 BIS-2X25
  dimensions (mm)
A   83.5 83.5 83.5 83.5 83.5 83.5
B   68.3 68.3 68.3 68.3 68.3 68.3
C   108.0 108.0 108.0 108.0 108.0 108.0
D   63.5 63.5 63.5 76.2 76.2 76.2
E   24.8 24.8 24.8 43.3 43.3 43.3
F   12.7 12.7 12.7 12.7 12.7 12.7
G   23.6 49.0 74.4 23.6 49.0 74.4
Fiberoptic
Rod O.D.
  19.1 19.1 19.1 19.1 19.1 19.1
  * note: The model number X descriptor determines the MCP option
  models * BIS-3X05 BIS-3X15 BIS-3X25 BIS-4X05 BIS-4X15 BIS-4X25
  dimensions (mm)
A   83.5 83.5 83.5 83.5 83.5 83.5
B   68.3 68.3 68.3 68.3 68.3 68.3
C   108.0 108.0 108.0 108.0 108.0 108.0
D   88.9 88.9 88.9 101.5 101.5 101.5
E   56.3 56.3 56.3 70.09 70.09 70.09
F   12.7 12.7 12.7 12.7 12.7 12.7
G   23.6 49.0 74.4 23.6 49.0 74.4
Fiberoptic Rod O.D.   19.1 19.1 19.1 19.1 19.1 19.1
  * note: The model number X descriptor determines the MCP option

BIS SPECIFICATIONS
imaging area:
9.1mm x 6.6 mm COLUTRON model BIS-1000 series,
18.2mm x 13.2 mm COLUTRON model BIS-2000 series,
27.3mm x 19.8mm COLUTRON model BIS-3000 series,
approx. 36.4mm x 26.4mm COLUTRON model BIS-4000 series
camera: CID
776H x 512V total pixels, 755H x 484V displayed pixels, 12.0µm x 13.7µm pixel size
+ 15V.dc nominal (+11.0 to + 17.0) input Voltage
RS-170, 2:1 interlace scanning format
Full output at 0.3fc faceplate illumination ( T=2,850°K )
Spectral Response: 0.11 Amps/Watt at l = 560 nm (P-20)
Quantum Efficiency: ~23 % at l = 560 nm (P-20)
Fiberoptic Faceplate 4.5µm pitch, 4mm thick
fiberoptic cable:
10.0 mm x 8.0 mm image area, 10 micron square fiber size
(Remote Systems) 600 mm length, flexible, UHV compatible
fiberoptic rod:
Available in 2", 3" or 4" lengths, 3/4" diameter
(Standard Systems) 6µm diameter fiber size
MCP (standard)
0.975" diameter, single plate (BIS-1000 series), chevron (BIS-1100 series)
1.289" diameter, single plate (BIS-2000 series), chevron (BIS-2100 series)
1.970" diameter, single plate (BIS-3000, BIS-4000 series), chevron (BIS-3100, BIS-4100 series)
10µm channel diameter
Max. gain: 2.0 x 10 4 (single plate), 1.0 x 10 7 (chevron)
Spatial resolution: 41 lp/mm
MSP (oPtional)
0.980" diameter, single plate (BIS-1200 series), chevron (BIS-1300 series)
1.291" diameter, single plate plate (BIS-2200 series), chevron (BIS-2300 series)
1.970" diameter, single plate (BIS-3200, BIS-4200 series), chevron (BIS-3300, BIS-4300 series)
50 micron sphere diameter
Max. gain: 1.5 x 10 6 (one plate, 3000V), 1.0 x 109 (dual plate, 4000V)
Spatial resolution: 2 lp/mm
Phosphor Screen:
P-20 deposited onto aluminized fiberoptic faceplate, fiber Size: 6µm diameter
Conversion efficiency: h = 0.063 photons / eV / electron
P-20 peak wavelength: l = 560 nm
Optional ITO (Indium Tin Oxide) coating available
Optional P-11, P-43, P-46, P-47 phosphor coatings available
Power Supply Requirements: ( see optional PCU power control units )
0 ~ +1,000 V, 1 mA single microchannel plate
0 ~ +2,000 V, 1 mA dual microchannel plate (chevron)
0 ~ +3,000 V, 1mA single microsphere plate
0 ~ +4,000 V, 1mA dual microsphere plate
0 ~ +5,000 V, 1 mA phosphor screen
Vacuum:
1 x 10-6 torr or better required to operate MCP, 1 x 10-5 or better to operate MSP
Good to at least 1 x 10-10 torr; maximum bakeout temp. 150°C

Colutron Beam Imaging products Local sales representative in Japan
APACE Science
Colutron Research Corporation APACE Science
2321 Yarmouth Avenue,
Boulder, Colorado 80301   U.S.A.
Mitomo-Bldg.#2 - 6F, Ebisu-nishi 2-Chome,
2 - 8, Shibuya-Ku, Tokyo 150-0021   Japan
phone#
Fax#
+1 (303) 443-5211
+1 (303) 449-5050
phone#
Fax#
+81 (0)3 - 3463 - 3204
+81 (0)3 - 3463 - 3205
contact: Erik Wahlin, Ph.D.
e-mail: erik@colutron.com
web: http://www.colutron.com/ ( Last updated: )

| Colutron Top-page | beam viewing systems-BVS | power-control units-PCU |
| image processing systems-IPS | phosphor screen | ion beam kits | APACE Science Home |